ProTest is a self-contained Biological Indicator for use in either
Steam or Ethylene Oxide sterilization. Each unit is made up of a plastic
container with cap, a crushable glass ampoule with recovery media, and a paper
strip inoculated with spores. ProTest is available for healthcare and
industrial use.
48-Hour Incubation
Suitable for use in a hospital or industrial Ethylene Oxide sterilization.
Contains B. atrophaeus1 spores
Log 6
Catalog # PT-1-6-25 (25 BIs)
Catalog # PT-1-6-100 (100 BIs)
Intended for use in EO 600mg/L, 60%RH, 54°C
Sterilization failure evident by a dramatic green to yellow color change
and/or turbidity.
|
|
 |
|
24-Hour Incubation
Intended for use in steam 121°C gravity and 121-135°C prevacuum cycles.
Suitable for use in a hospital or industrial Steam Sterilization.
Contains G. stearothermophilus1 spores
Log 5
Catalog # PT-3-5-25 (25 BIs)
Catalog # PT-3-5-100 (100 BIs)
Log 6
Catalog # PT-3-6-25 (25 BIs)
Catalog # PT-3-6-100 (100 BIs)
|
The control unit should exhibit turbidity and/or a color change to
or toward yellow. If the control unit does not show signs of growth,
consider the test invalid.
A failed sterilization cycle is indicated by turbidity and/or a
color change to or toward yellow. A test unit that retains its original
color (green for ProTest EO and purple for ProTest Steam) indicates that
sterilization parameters have been met.
Each lot of ProTest is certified for population, species, D-value,
z-value (where applicable), purity and expiry. Shelf life is 18 months from the
date of manufacture. The above catalog numbers for ProTest comply with the
AAMI ST19/EN 866-7 (Steam) and AAMI ST24/EN 866-8 (Ethylene Oxide).
Note:
- Derived from recognized reference strains as specified within USP, ISO or
EN guidelines. These products are not certified as being representative of
the reference strain and they should therefore not be used where a reference
culture is specified.
|
|